AD-Ⅱ Neonatal CPAP system

Contact Us
Online Message
Detail

AD-Ⅱ CPAP   system for infants/neonates

CPAP system is a neonatal respiratory support medical device designed based on the wide application of non-invasive continuous positive airway pressure in the pediatric field. It is designed for premature infants, newborns, infants and children up to 14 years old. It is used for nasal congestion CPAP ventilation treatment. CPAP mode keeps children breathing spontaneously early, and can effectively reduce children's work of breathing.


Superiority

/The system structure is compact, and the overall design of the host, medical electric humidifier(optional), and medical air compressor (optional);

/Built-in air-oxygen mixer, the oxygen concentration is continuously adjustable, no need to calculate or refer to the comparison table;

/The flow rate and oxygen concentration are adjusted separately and independently, and they do not affect each other;

/Real-time monitoring of oxygen concentration, double guarantee the safety of oxygen use;

/CPAP regulates the pressure, combined with system pressure monitoring and safety valves to fully ensure the safety of patients;

/ The host is a passive product, no power supply is needed, and it is equipped with a universal mechanical boom;

/Optional medical air compressor for drying, low noise;

/Optional electronically controlled heated humidifier, the system to ensure good wet state; light conduit and nasal congestion, serve to reduce the resistance of the breathing circuit, to reduce the work of breathing action of the patient fixing device with the system, reduces the compression of the patient's head and face and damage to the nasal septum;

/Suitable for pediatric PIICU , NICU , PICU and emergency room.

technical parameter:

Ventilation mode (MODE): CPAP mode

Oxygen concentration adjustment: 21% ~ 100%

Flow adjustment: 0 ~ 18LPM

Continuous positive airway pressure adjustment ( CPAP adjustment): 0 ~ 1.0kPa (0 ~1 0cmH2O)

monitor:

Pressure monitoring: 0 ~ 2.5kPa (0 ~ 25 cmH2O)

Pressure safety valve: When the pressure in the equipment pipeline exceeds 1.5 times the adjustment parameter , the safety valve opens automatically.

Alarming:

Differential pressure alarm: when the air supply pressure difference is greater than 0.1MPa ,Sound alarm

General parameters:

Host size: W × H × D 25cm × 25.5cm × 21cm.

Host weight: about 4KG

Air supply: 0.3MPa ~ 0.4MPa